Utilities
SMIQ
1125.5555.03
E-9
2.418
Calibration VCO SUM
To synchronize the summing loop, the frequency the oscillator generates must be so close to the rated
frequency that the phase control can lock in. This is effected by means of presetting values. The
presetting values are stored in a table and can be renewed using internal calibration routine VCO SUM.
The calibration routine needs only be executed after a data loss in the RAM or after an exchange of
modules.
Menu selection:
UTILITIES - CALIB - VCO SUM
SYST EM
REF OSC
PHAS E
PROT ECT
CALI B
DIAG
TEST
MOD KEY
AUX I/O
BEEP ER
INST ALL
ALL
VCO SUM
VECT OR MOD
LEV PRESET
OSC REF
LEVE L
ALC TABLE
LEV ATT
FREQ
100. 000 000 0
MHz
LEVE L
- 30.0
dBm
CALIBRATE
VIEW
750.0 MHz
760.0 MHz
770.0 MHz
780.0 MHz
790.0 MHz
800.0 MHz
810.0 MHz
820.0 MHz
2
2
2
2
2
2
2
2
0001
0002
0003
0004
0005
0006
0007
0008
-22mV
-24mV
-25mV
-26mV
-27mV
-28mV
-29mV
-30mV
88
108
127
147
167
187
207
217
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-262
Menu UTILITIES - CALIB - VCO SUM
CALIBRATE
Triggers the calibration for the VCO summing loop.
IEC/IEEE-bus command
:CAL:VSUM?
VIEW
Indication of the list of correction values.
The cursor wraps to index 1 of the list. The list can be executed using the rotary
knob. This index can be obtained by entering the index value on the digit block.
IEC/IEEE-bus command
:CAL:VSUM:OFFS?
:CAL:VSUM:DAC?
:CAL:VSUM:KOS?