Digital Standard NADC
SMIQ
1125.5555.03
E-9
2.278
(SELECT SLOT...)
SHORT
(UPLINK ONLY)
SLOT LEVEL
Selection of level for selected slot.
OFF
Maximum attenuation
IEC/IEEE
:SOUR:NADC:SLOT2:LEV OFF
FULL
The level corresponds to the value indicated on
the SMIQ LEVEL display.
IEC/IEEE
:SOUR:NADC:SLOT2:LEV FULL
ATTEN
The level is reduced by the value set under
SLOT ATTENUATION.
IEC/IEEE
:SOUR:NADC:SLOT2:LEV ATT
SET DEFAULT
Resets the subsequent parameters to the factory-set values.
IEC/IEEE-bus command
:SOUR:NADC:SLOT8 PRES
G
Display of data contents in the 6 bit data field "Guard" in
hexadecimal form.
R
Display of data contents in the 6 bit data field "Ramp" in
hexadecimal form..
SYNC
Input value for the sync word in hexadecimal form.
Depending on the selected slot, the following sync words are
predefined according to the NADC standard:
SLOT 1
A91DE4A
SLOT 2
A9D127A
SLOT 3
C7E3C0C
SLOT 4
342C3F3
SLOT 5
13E23D1
SLOT 6
DC2EC1D
IEC/IEEE-bus
:SOUR:NADC:SLOT2:SYNC A91EE4A
CDVCC
Input value for the 12 bit "Coded digital verification color
code" field in hexadecimal form. SMIQ does not perform any
coding but directly accepts the input data. 013 hex. is preset.
This corresponds to a value of 1 for the uncoded colour code
DVCC.
IEC/IEEE-bus
:SOUR:NADC:SLOT2:CDVC #HFFF
V, W, X, Y
Display of data contens of other data fields that are
predefined by the NADC standard.
G2
Display of data contents in the second "Guard" in
hexadecimal form.
Parameter
SELECT DATA LIST...
to
EDIT DATA LIST...
see menu selection BURST TYPE = TCH