SMIQ
Digital Standard IS-95 CDMA
1125.5555.03
E-9
2.149
CHANNEL
TYPE/RATE
Selects the type of channel and the associated data rate. This selection is only
possible in the reverse link mode. The selection also determines the structure
of the channel coding and the number of data bits to be inserted into each
frame.
TRAFFIC 14400
Traffic channel with 14400 bps. 267 bits are inserted into
each frame
TRAFFIC 7200
Traffic channel with 7200 bps. 125 bits are inserted into
each frame
TRAFFIC 3600
Traffic channel with 3600 bps. 55 bits are inserted into
each frame
TRAFFIC 1800
Traffic channel with 1800 bps. 21 bits are inserted into
each frame
ACCESS 4800
Access channel with 4800 bps. 88 bits are inserted into
each frame
TRAFFIC 9600
Traffic channel with 9600 bps. 172 bits are inserted into
each frame
TRAFFIC 4800
Traffic channel with 4800 bps. 80 bits are inserted into
each frame
TRAFFIC 2400
Traffic channel with 2400 bps. 40 bits are inserted into
each frame
TRAFFIC 1200
Traffic channel with 1200 bps. 16 bits are inserted into
each frame
IEC/IEEE-bus
:SOUR:IS95:RLC:CTYP TRAF7200
:SOUR:IS95:RLC:CTYP ACC4800
DATA SOURCE
Selects the modulation data for the reverse link signal with channel coding. The
data are continuously inserted into the data field of the corresponding frame.
PN..
PRBS data to CCITT with period length between 2
9
-1 and 2
23
-1.
IEC/IEEE-bus command
:SOUR:IS95:RLC:DATA PN1
DLIST
Data from a previously programmed and stored data list.
IEC/IEEE-bus command
:SOUR:IS95:RLC:DATA DLISt
ADD FRAME QUAL
INDICATOR
Switches the frame quality indicator (CRC calculation) on or off. With OFF
selected, only zeros are sent instead of CRC bits.
IEC/IEEE-bus command
:SOUR:IS95:RLC:FQIN ON
CONVOLUTIONAL
ENCODER
Switches the convolutional encoder on or off. With OFF selected, the data rate
is attained by repeating the symbol.
IEC/IEEE-bus command
:SOUR:IS95:RLC:CENC ON
BLOCK
INTERLEAVER
Switches the interleaver function on or off.
IEC/IEEE-bus command
:SOUR:IS95:RLC:BINT ON
ERASURE BIT
Selects 1 or 0 for the erasure bit.
IEC/IEEE-bus command
:SOUR:IS95:RLC:EBIT 0
Parameter
SELECT DATA LIST...
to
EDIT DATA LIST...
see section "Menu IS-95 CDMA Standard -
Forward Link Signal"