Sweep
SMIQ
1125.5555.03
2.392
E-9
2.26 Sweep
The SMIQ offers a digital step-by-step sweep for parameters:
•
RF frequency
•
LF frequency
•
RF level
Setting a sweep is effected in five basic steps which are shown in the following example, the setting of a
frequency sweep:
1. Set sweep range (START and STOP or CENTER and SPAN).
2. Select linear or logarithmic sequence (SPACING).
3. Set step width (STEP) and dwell time (DWELL).
4. Activate marker if desired (MARKER).
5. Switch on sweep (MODE set to AUTO, SINGLE or STEP).
2.26.1
Setting the Sweep Range (START, STOP, CENTER and SPAN)
The sweep range of the RF sweep can be entered in two different ways. Either by entering the START
and STOP value or by entering CENTER and SPAN. Please observe that the two parameter sets
influence one another. The influence is exerted in the following way:
START frequency altered:
STOP
=
unaltered
CENTER =
(START + STOP)/2
SPAN
=
(STOP - START)
STOP frequency altered:
START
=
unaltered
CENTER =
(START + STOP)/2
SPAN
=
(STOP - START)
CENTER frequency altered
SPAN
=
unaltered
START
=
(CENTER - SPAN/2)
STOP
=
( SPAN/2)
CENTER =
unaltered
START
=
(CENTER - SPAN/2)
STOP
=
( SPAN/2)