SMIQ
Digital Standard GSM/EDGE
1125.5555.03
E-9
2.311
(SAVE/RCL FRAME...) GET PREDEFINED
FRAME...
Loads a factory-set frame configuration.
NB0
Normal Burst in slot 0
IEC/IEEE-bus
:SOUR:GSM:FLIS:PRED:LOAD "nb0"
RECALL FRAME...
Loads a frame configuration saved by the user.
IEC/IEEE-bus
:SOUR:GSM:FLIS:LOAD "name"
SAVE FRAME...
Saves a user-defined frame configuration.
IEC/IEEE-bus
:SOUR:GSM:FLIS:STOR "name"
DELETE FRAME...
Deletes a frame configuration saved by the user.
IEC/IEEE-bus
:SOUR:GSM:FLIS:DEL "name"
SELECT SLOT...
Selection of one of 8 possible slots. When selecting the slot, a window is
opened in which the data contents belonging to this slot can be defined.
If the cursor is placed onto a slot in the diagram, it may be switched on and off
by pressing one of the unit keys (toggle function).
Menu selection:
BURST TYPE = NORM
Fig. 2-201
Menu DIGITAL STD - GSM/EDGE - SELECT SLOT - NORM, SMIQ equipped with
Modulation Coder SMIQB20 and Data Generator SMIQB11