SMIQ
Digital Standard IS-95 CDMA
1125.5555.03
E-9
2.139
REV LINK FILTER...
Opens a window for selecting the baseband filter for
the reverse link. A selection between the FIR filters
defined in IS-95, the Nyquist filters COS and
SQRCOS or a user-defined filter USER (cf. Section
Digital Modulation) is possible. A FIR filter is preset
according to IS-95.
IEEE command
:SOUR:IS95:FILT:RTYP IS95
ROLL OFF FACTOR
Setting value of roll-off factor.
The setting is only possible provided that a COS or
SQRCOS filter is selected under FILTER.
IEEE-bus command
:SOUR:IS95:FILT:PAR 0.22
FILTER MODE
Selection of filter mode.
LOW_ACP
Filter for minimum Adjacent Channel
Power.
IEEE
SOUR:IS95:FILT:MODE LACP
LOW_EVM
Filter for minimum vector error.
IEEE
SOUR:IS95:FILT:MODE LEVM