RF Level
SMIQ
1125.5555.03
E-8
2.52
2.5.4
User Correction (UCOR)
Function "User Correction" can be used to create and activate lists in which arbitrary RF frequencies are
assigned level correction values.
Up to 10 lists with a total of 160 correction values can be compiled. For frequencies which are not
included in the list the level correction is determined by means of interpolation of the nearest correction
values.
When user correction is switched on, the LEVEL indication is completed by the indication UCOR (User
Correction) in the header field of the display. The RF output level is the sum of both values.
LEVEL + UCOR = output level
If the offset setting is used at the same time, the LEVEL indication value is the difference of the input
values AMPLITUDE and OFFSET of the menu LEVEL.
AMPLITUDE - OFFSET = LEVEL
The user correction is effective in all operating modes if switched on.
Menu selection:
LEVEL - UCOR
LEVEL
ALC
UCOR
EMF
STATE
SELECT LIST...
CURRENT: UCOR1
DELETE LIST...
FUNCTION
FILL
INSERT
DELETE
EDIT/VIEW
FREQ
100. 000 000 0
MHz
LEVEL
- 27.0
dBm
UCOR
+ 1.9
dB
OFF
ON
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-21
Menu LEVEL - UCOR - OPERATION side
STATE
Switching on/off user correction.
IEC/IEEE-bus command
SOUR:CORR ON
SELECT LIST...
Selection of a list or generation of a new list (cf. Section 2.2.11, List Editor).
IEC/IEEE-bus command
SOUR:CORR:CSET "UCOR1"
DELETE LIST...
Deletion of a list (cf. Section 2.2.11, List Editor).
IEC/IEEE-bus command
SOUR:CORR:CSET:DEL "UCOR2"
FUNCTION
Selection of the editing mode to process the selected list (cf. Section 2.2.11, List
Editor).
IEC-bus commands
SOUR:CORR:CSET:DATA:FREQ 100 MHz, 102 MHz,...
SOUR:CORR:CSET:DATA:POW 1dB,
0.8dB,...