SMIQ
LIST Mode
1125.5555.03
2.403
E-9
MARKER
At the first step of the LIST mode, this output provides an approx. 200 µs
trigger signal immediately after blanking. At small DWELL times, this signal
can be used for an accurate synchronization to trigger other devices and
shows the first stable output frequency. The delay to the fed-in signal at the
TRIGGER input for EXT-SINGLE or EXT-STEP is 1.5 to 2 ms and has a jitter
of 0.5 ms.
TRIGGER
Input
BLANK
Output
MARKER
Output
Frequency
Fig. 2-249
Signal example LIST mode: MODE = EXT-STEP
The LIST menu offers access to settings for the LIST mode.
Menu selection:
LIST
LEVE L
- - -.-
- - -.- - -
MODE
OFF
AUTO
SIN GLE
STEP
EXT-SI NGLE
EXT-ST EP
RESE T LIST
DWEL L
10.0 ms
CURR ENT IN DEX
1
LEAR N
SELE CT LIS T...
C URRENT : LIST 2
DELE TE LIS T...
FUNC TION
FILL
INSERT
DEL ETE
EDIT/VI EW
FREQ
LIST
EXECUTE SINGLE LIST
FREQ UENCY
LEVE L
ANAL OG MOD
VECT OR MOD
DIGI TAL MO D
DIGI TAL ST D
LF OUTPUT
SWEE P
LIST
MEM SEQ
UTIL ITIES
Fig. 2-250
Menu LIST - OPERATION page