SMIQ
List Editor
1125.5555.03
2.35
E-8
2.2.11.3 Edition of Lists
Due to the selection of an edit mode on the OPERATION page the EDIT page is automatically activated.
When the EDIT/VIEW function is selected, the largest possible section of the list is displayed (cf. Fig.
2-10). In the case of block functions FILL, INSERT and DELETE, an input window is additionally
displayed (cf. Fig. 2-11 to 2-13).
Functions SELECT LIST and FUNCTION are available on the EDIT page as on the OPERATION page.
Return to the OPERATION page is effected by pressing the [SELECT] key twice.
Single-value function EDIT/VIEW
By selecting the EDIT/VIEW function, the entire list can be viewed or modifications of single values be
carried out.
If the cursor marks a value in the INDEX column of the list, the EDIT mode is exited by pressing the
[RETURN] key. The menu cursor then marks FUNCTION again.
There is no separate function for storing the list. This means that every modification of the list is
transferred to the internal data set and has an effect on exiting the EDIT/VIEW function.
Menu selection:
FUNCTION EDIT/VIEW
DELETE
FREQ
100. 000 000 0
MHz
L E V E L
-30.0
d B m
INSERT
FILL
EDIT/VIEW
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
S E L E C T L I S T . . .
C U R R E N T : M S E Q 2
F U N C T I O N
- I N D E X - F R E E 2 4 6
- L E N 1 0
M E M O R Y
D W E L L
0 0 0 1
0 9
5 0 m s
0 0 0 2
0 2
5 0 m s
0 0 0 3
0 1
6 0 m s
0 0 0 4
2 3
6 0 m s
0 0 0 5
0 9
8 5 m s
0 0 0 6
1 0
8 5 m s
0 0 0 7
0 8
8 5 m s
0 0 0 8
1 1
8 5 m s
Fig. 2-10
Edit function EDIT/VIEW
INDEX
Position in the list
FREE
Space available. FREE 256 means that space for 256 parameter elements is
available in the list memory in total.
LEN
Occupied space. LEN 0 means that the current list occupies 0 elements in the
list memory.
MEMORY DWELL
Identification of the column below. The number and name of parameter
columns is different for the various list editors.