Digital Modulation
SMIQ
1125.5555.03
E-9
2.110
(TRIGGER...)
TRIGGER SOURCE
Selection of trigger source
EXT
An external trigger signal can be fed in at
TRIGIN of connector PAR DATA at the rear
of SMIQ. With the active edge, a data
sequence is started from the data generator
memory.
INT
With INT selected, a trigger event can be
manually executed by EXECUTE TRIGGER.
IEC/IEEE-bus
:SOUR:DM:TRIG:SOUR EXT
EXT TRIGGER
DELAY
Input value of number of symbols by which an
external trigger signal is delayed before it starts the
data sequence in the data generator. A synchroneity
with the DUT or other units can thus be achieved.
IEC/IEEE-bus
:SOUR:DM:TRIG:DEL 3
EXT RETRIGGER
INHIBIT
Input of number of symbols for which each new
trigger event is inhibited during MODE RETRIG after
a trigger signal.
During MODE RETRIG, each new trigger signal
restarts the data sequence in the data generator.
This restart can be inhibited for the entered number
of symbols. The entry of 1250 symbols, for example,
causes new trigger signals to be ignored for the
duration of 1250 symbols after execution of a trigger
event.
IEC/IEEE-bus
:SOUR:DM:TRIG:INH 1250
CLOCK...
Opens a window for selecting the clock source and for setting a delay.
FREQ
100. 000 000 0
MHz
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Π
ΠΠ
Π
/4DQPSK
CLOCK SOURCE
INT
COUPLED
EXT
MODE
SYMBOL
BIT
DELAY
0.00 Symb
LEVEL
- 30.0
dBm
PEP
- 27.6
dBm
STATE
SOURCE...
SELECT STAN
MODULATION.
SYMBOL RATE
FILTER...
CODING...
TRIGGER MOD
EXECUTE TRI
TRIGGER...
CLOCK...
Fig. 2-66
DIGITAL MOD - CLOCK, SMIQ equipped with option Modulation Coder SMIQB20 and
option Data Generator SMIQB11