SMIQ
Utilities
1125.5555.03
2.417
E-9
2.29.9 Calibration
(CALIB)
For servicing, the following menus offer access to calibrating routines and correction values:
UTILITIES - CALIB -
ALL
VCO SUM
VECTOR MOD
LEV PRESET
REF OSC
(cf. service manual)
LEVEL
(cf. service manual)
ALC TABLE
LEV ATT
LFGEN
Internal calibration routines LEV PRESET and VCO SUM are protected by a password. They can only
be executed if the lock-out in the UTILITIES - PROTECT menu has been unlocked. The password is
PASSWORD LEVEL 1 = "123456".
Caution:
Execute calibration routines only when the instrument has warmed up
Calibration routines LEVEL and REF OSC are described in the service manual.
Calibration ALL
CALIB ALL triggers all internal calibrations which do not require any external measuring equipment. The
calibrations with external measurements are described in the service manual.
Menu selection:
UTILITIES - CALIB - ALL
SYSTEM
REF OSC
PHASE
PROTECT
CALIB
DIAG
TEST
MOD KEY
AUX I/O
BEEPER
INSTALL
ALL
VCO SUM
VECTOR MOD
LEV PRESET
OSC REF
LEVEL
ALC TABLE
LEV ATT
LFGEN
FREQ
100. 000 000 0
MHz
LEVEL
- 30.0
dBm
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
CALIBRATE ALL
Fig. 2-261
Menu UTILITIES - CALIB - ALL
CALIBRATE ALL
Triggers all internal calibrations.
IEC\IEEE-bus command
:CAL:ALL?