SMIQ
RF Level
1125.5555.03
E-8
2.53
Menu selection:
LEVEL - UCOR
LEVEL
ALC
UCOR
EMF
FREQ
100. 000 000 0
MHz
LEVEL
- 27.0
dBm
UCOR
+ 1 .9
dB
SELECT LIST...
CURRENT: UCOR1
FUNCTION
FILL
INSERT
DELETE
EDIT/VIEW
INDEX - FREE 70 - LEN 10
FREQUENCY
UCOR1
005
105.000 000 0 MHz
+ 1.9 dB
006
107.000 000 0 MHz
+ 1.2 dB
007
108.000 000 0 MHz
+ 1.3 dB
008
109.000 000 0 MHz
+ 1.5 dB
009
111.000 000 0 MHz
+ 1.6 dB
010
112.000 000 0 MHz
+ 1.9 dB
011
113.000 000 0 MHz
+ 2.0 dB
012
114.000 000 0 MHz
+ 2.1 dB
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-22
Menu UCOR - LEVEL-EDIT side
2.5.5 EMF
The signal level can also be set and indicated as the voltage of EMF (open-circuit voltage).
EMF is displayed in the header field of the display after the unit of the level indication.
Menu selection:
LEVEL - EMF
LEVEL
ALC
UCOR
EMF
STATE
FREQ
100. 000 000 0
MHz
LEVEL
14.1
mV
OFF
ON
emf
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
Fig. 2-23
Menu LEVEL-EMF
STATE
ON
Voltage value of the level is the voltage of EMF.
OFF
Voltage value of the level is voltage at 50
Ω
(preset setting).