SMIQ
Digital Standard PDC
1125.5555.03
E-9
2.297
(SELECT SLOT...)
SYNC
SET DEFAULT
Resets the subsequent parameters to the factory-set values.
IEC/IEEE-bus command
:SOUR:PDC:SLOT8 PRES
G
(UPLINK only)
Display of data contents in the 54 bit data field "Guard" in
hexadecimal form.
The field is only displayed with link direction UPLINK
selected.
R
Display of data contents in the 6 bit data field "Ramp".
P
Input value of the data field "Preamble" in hexadecimal
form.The length of the field is 48 bit with link direction =
UPLINK, and 102 bit with DOWNLINK
IEC/IEEE-bus command
:SOUR:PDC:SLOT2:PRE #HFF
SYNC
Input value for the 32 bit sync word in hexadecimal form.
Depending on the selected slot and the link direction, the
following sync words are predefined according to the PDC
standard:
Slot No.
Downlink Sync Word
Uplink Sync Word
0
2F94D06B
D06B2F94
1
1D4EE2B1
E2B11D4E
2
70168FE9
8FE97016
3
83527CAD
7CAD8352
4
3678C987
C9873678
5
48D8B727
B72748D8
According to the standard, the sync words of the uplink are
the inverted sync words of the downlink.
IEC/IEEE-bus
:SOUR:PDC:SLOT2:SYNC #HE2B11D4F
SI
Selection of data source for "Sync Information" field. The
data in each of the three SI fields are identical
PN..
PRBS data according to CCITT with periods
between 2
9
-1 and 2
23
-1.
IEC/IEEE
:SOUR:PDC:SLOT2:SOUR:SI PN15
DLIST
Data from a programmable data list.
IEC/IEEE
:SOUR:PDC:SLOT2:SOUR:SI DLIS
SERDATA Data from data input SER DATA.
IEC/IEEE
:SOUR:PDC:SLOT2:SOUR:SI SDAT
Q
Display of the tail bits.
PO
(DOWNLINK only)
Input value for the 78 bit "Postamble" field in hexadecimal
form.
IEC/IEEE-bus command
:SOUR:PDC:SLOT2:POST #HFF
G2
(UPLINK only)
Display of data contents in the 78 bit "Guard" field in
hexadecimal form at the end of the slot.
Parameter
SELECT DATA LIST...
to
EDIT DATA LIST...
see menu selection BURST TYPE = TCH