Signal Generator Settings
R&S
®
SGT100A
118
User Manual 1176.8674.02 ─ 07
"Next Segment Seamless"
The segment selected under "Next Segment" is output.
This mode is only available if all segments have the same sample
rate.
●
In the case of "Internal Trigger = Auto", output starts at once and
the segment is generated continuously.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "internal Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
If the segment is changed in"Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
●
In the case of "External Trigger = Auto", output starts at once and
the segment is generated continuously.
If the segment is changed in "Next Segment", the new segment is
output seamlessly after the output of the current segment is com-
plete.
●
In the case of "External Trigger = Armed_Auto", output starts after
the first trigger event. The segment is then generated continu-
ously.
If the segment is changed in "Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
The remaining trigger modes ("Retrigger" and "Armed_Retrigger") are
not available.
"Sequencer"
The waveform files are processed according to the order and repeti-
tion cycles defined in the special sequencing list file (
*.wvs
)
assigned to the multi segment waveform file.
This “play list” is defined in the
Multi Segment Waveform Sequencing
dialog.
Note:
This feature is disabled in case the segments in the multi seg-
ment waveform file have different clock rates.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe
Execute Next Segment
Triggers manually switchover to the subsequent segment in the multi segment file. A
manual trigger can be executed only when an internal "Next Segment Source" has
been selected.
To perform a switchover to any segment within the multi segment file, adjust the
parameter "Next Segment".
This parameter is disabled, if a sequencing play list is enabled.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT:EXECute