Remote Control Commands
R&S
®
SGT100A
327
User Manual 1176.8674.02 ─ 07
:PFASt
<Pow>
Special command to set the RF output level with minimum latency at the RF output
connector. This value does not consider a specified offset. No unit (e.g. dBm) allowed.
Bypasses the status system so command
*OPC?
cannot be appended.
Parameters:
<Pow>
float
Example:
:PFASt -20
11.11 OUTPut Subsystem
...................................................................................327
....................................................................................327
:OUTPut:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set without the attenuator being adjusted
(Attenuator FIXed).
Return values:
<Lower>
float
Usage:
Query only
Manual operation:
See
:OUTPut:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set without the attenuator being adjusted
(Attenuator FIXed).
Return values:
<Upper>
float
Usage:
Query only
Manual operation:
See
:OUTPut:AMODe
<AMode>
Switches the mode of the attenuator at the RF output.
OUTPut Subsystem