General Instrument Settings and Instrument Setup
R&S
®
SGT100A
245
User Manual 1176.8674.02 ─ 07
I/Q Modulator
Starts the adjustment for the I/Q modulator for the entire frequency range. The I/Q
modulator is adjusted regarding carrier leakage, I/Q imbalance and quadrature.
Remote command:
:CALibration:IQModulator:FULL?
8.2 Hardware Configuration
Querying information about the installed assemblies
► Select "SGMA-GUI > Instrument Name > Setup > Hardware Config".
In the "Hardware Config" dialog, the installed assemblies together with their var-
iants and revision states can be displayed for servicing purposes.
The dialog is a table that lists the installed assemblies. It is divided into the sec-
tions:
● "Common Assembly"
● "RF Assembly"
Assembly
The tables list the installed assemblies.
Hardware Configuration