Remote Control Commands
R&S
®
SGT100A
469
User Manual 1176.8674.02 ─ 07
Parameters:
<Amplitude>
float
Range:
-120 to 25
Increment: 0.01
*RST:
-10
Default unit: dBm
Manual operation:
See
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet
<Offset>
Specifies the constant level offset of a downstream attenuator/amplifier. If a level offset
is entered, the level entered with
:POWer
no longer corresponds to the RF output
level.
The following correlation applies:
:POWer
= RF output level +
POWer:OFFSet
.
Entering a level offset does not change the RF output level, but rather the query value
of
:POWer
.
Only dB is permitted as the unit here. The linear units (V, W, etc.) are not permitted.
Parameters:
<Offset>
float
Range:
-100 to 100
Increment: 0.1
*RST:
0
Manual operation:
See
[:SOURce]:POWer:RANGe:LOWer?
[:SOURce]:POWer:RANGe:UPPer?
Queries the minimum/maximum level range in the current level mode
Return values:
<Upper>
float
Usage:
Query only
Manual operation:
See
[:SOURce]:POWer:WIGNore
<State>
Ignores level range warnings.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
POW:WIGN ON
suppresses the level range warnings.
SOURce Subsystem