Remote Control Commands
R&S
®
SGT100A
460
User Manual 1176.8674.02 ─ 07
<DeltaPower>
float
Range:
-20 to 20
Increment: 0.01
*RST:
0
Example:
See
Example "Defining correction coefficients and enabling digi-
Usage:
Query only
11.14.15 SOURce:PHASe Subsystem
........................................................................................ 460
[:SOURce]:PHASe
<Phase>
Specifies the phase variation relative to the current phase.
Parameters:
<Phase>
float
Range:
-360 to 360
Increment: 0.1
*RST:
0
Default unit: DEG
Manual operation:
See
[:SOURce]:PHASe:REFerence
Adopts the phase set with command
as the current phase.
Usage:
Event
Manual operation:
See
11.14.16 SOURce:PULM Subsystem
This subsystem contains the commands for setting the pulse modulation.
Programming Examples
Example: Performing pulse modulation
This example shows a command sequence to perform pulse modulation.
// ******************************************************************
// Reset the instrument to start from an initial state
// ******************************************************************
*RST; *CLS
SOURce Subsystem