Remote Control Commands
R&S
®
SGT100A
359
User Manual 1176.8674.02 ─ 07
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLUNit
<SLUnit>
The command defines the unit for the entry of the length of the signal sequence
(
SOUR:BB:ARB:TRIG:SLEN
) to be output in the "Single" trigger mode
(
SOUR:BB:ARB:SEQ SING
).
Parameters:
<SLUnit>
SEQuence | SAMPle
*RST:
SEQuence
Example:
BB:ARB:SEQ SING
sets trigger mode Single.
BB:ARB:TRIG:SLUN SEQ
sets unit Sequence length for the entry of sequence length.
BB:ARB:TRIG:SLEN 2
sets a sequence length of 2 waveforms. The current waveforms
will be output twice after the next trigger event.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe
<SMode>
The command selects the extended trigger mode for multi segment waveforms.
Parameters:
<SMode>
SAME | NEXT | NSEam | SEQuencer
NSEam = Next Segment Seamless
*RST:
NEXT
Example:
SOUR:BB:ARB:SEQ AUTO
selects trigger mode AUTO.
SOUR:BB:ARB:TRIG:SMOD SAME
the same segment of the waveform is output repeatedly.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SOURce
<Source>
Selects the trigger source.
Parameters:
<Source>
INTernal|OBASeband|BEXTernal|EXTernal
INTernal
manual trigger or
*TRG
.
EXTernal|BEXTernal
trigger signal on the USER 1/2 connector.
*RST:
INTernal
Example:
SOURce1:BB:ARBitrary:TRIGger:SOURce EXTernal
sets external triggering via the USER 1/2 connector.
Manual operation:
See
SOURce Subsystem