- 76 -
Means of ingredient verification:
Use this function to specify how to determine ingredient mass during formulation making process,
by TOLERANCE or by THRESHOLDS.
•
TOLERANCE: the software accepts mass of an ingredient as correct if it is within set percent
tolerance of total ingredient’s mass (±%) - (data acquired from database of products).
•
THRESHOLDS: the software accepts mass of an ingredient as correct if it is within set
thresholds (data acquired from database of products).
Specified method is in effect for all ingredients in the formula making process.
If during formulation making process one of the ingredients is weighed in excess (its mass is
out of high tolerance limit) the following message is displayed upon ingredient weight
confirmation: <VALUE OUT OF RANGE, RECALCULATE THE FORMULATION?>. Confirm,
the software automatically recalculates mass of other ingredients so that the mixture
proportions are correct.
In order to make THRESHOLDS option active, the product/ingredient related data in database
must correspond
with value selected for <MEANS OF INGREDIENT VERIFICATION>
parameter, i.e. ingredient thresholds are specified and <THRESHOLDS> parameter is selected
for <MEANS OF INGREDIENT VERIFICATION> parameter, otherwise particular ingredients
mass recalculation is not possible, i.e. ingredient thresholds are specified and <TOLERANCE>
parameter is selected for <MEANS OF INGREDIENT VERIFICATION> parameter.
Multiplier’s editing
Function designed to enable making multiple mixtures according to a selected formulation in a single
weighing process:
•
YES - on selecting a formulation for processing the software requests for multiplier’s value, the
provided value is in effect for all ingredients in the formulation making process, i.e. each
ingredient is multiplied by the value.
•
NO - multiplier value is disabled, default value is set to <1>
Printout,
for information on how to operate this function read section 15.7 “Additional Parameters for
a Weighing Process”.
22.2. QUICK ACCESS KEYS
Each working mode features automatically displayed set of default keys. The set can be modified by
assigning out-of-list quick access keys to on-screen keys. Such operation requires particular
permissions level.
22.3. ADDING FORMULATION TO FORMULATIONS DATABASE
Formulation database contains formulations names, ingredients names and their weight. Each
product used for formulation making process is saved to the database of products. When preparing a
formulation, first name it, next add ingredients.
The software is intuitive and you are guided by means of respective prompts. Since each ingredient
must have its name and mass specified you have to know the exact composition of the total mixture.
In order to add a new formulation enter the formulation database.
Procedure:
1. Enter <
Databases> submenu, press <Formulations> button.
2. Press <
Add> button for adding a new formulation.
New record is added and edited automatically. Enter necessary data.
List of formulation parameters:
•
Name: click name button, new window opens, go to the window and enter formulation name.
•
Code: click code button, enter code number.
•
Ingredients: click ingredients button, list of formulation ingredients is displayed (for new
formulation the list is empty), add the ingredients:
Summary of Contents for UYA 4Y
Page 8: ...8...
Page 9: ...9 1 GENERAL INFORMATION 1 1 DIMENSIONS Ultra microbalances and microbalances...
Page 10: ...10 MYB 4Y Microbalances Indicator XA 4Y series balances...
Page 11: ...11 XA 4Y F series balances...
Page 195: ...195...
Page 196: ...196...