- 49 -
as the indication returns to autozero zone, the software automatically deletes tare
value.
Automatic footer printout
Options:
•
MODE –
No
– manual footer printout,
Sum of measurements
– footer is printed when mass value entered in
<Threshold> parameter is out of range
Measurements quantity
– footer is printed when a pre-defined
measurements quantity (a batch) set in parameter <Threshold> is carried
out
•
THRESHOLD – setting the threshold value determining footer printout.
For <Sum of measurements> parameter threshold is determined in
measuring unit [g], for <Measurements quantity> parameter the unit is not
determined, provided value specifies carried out measurements quantity.
Printout mode / release
•
PRINT / RELEASE button, (manual operation)
Never
– printout deactivated
First stable
– the first stable measurement is printed
Each stable
– all stable measurements are set for printing
Each
– all measurements are printed (stable and unstable), in case of a
verified balance only stable ones (as in setting <Each stable>)
•
AUTOMATIC MODE
Never
– printout deactivated
First stable
– the first stable measurement result obtained after placing a
load on balance’s weighing pan is recorded, record of the following
measurement result takes place only on unloading the weighing pan,
returning of the mass indication below set threshold’s value, and placing
another load on the weighing pan
Last stable
– only the last stable measurement is accepted, i.e.
measurement recorded before taking the load off the weighing pan. The
record takes place on taking the load off the weighing pan, and returning of
the mass indication below set threshold’s value.
With interval
– automatic printout and record of weighing results, carried
out in a cyclic manner as often as specified by <INTERVAL> parameter.
The interval is set in minutes, values range is 1 – 9999.
CAUTION!
Each weighing result is printed and recorded (both stable and unstable for
non-verified balance, stable for verified balance).
The measurements start to be printed and recorded as soon as you enable
automatic operation with interval. First stable weighing result of value higher
than THRESHOLD value is recorded as the first measurement. Successive
measurements are printend as often as specified by <INTERVAL> parameter.
To end automatic operation with interval disable this function.
•
THRESHOLD
mass value obligatory for operation of automatic printout, set in grams.
•
INTERVAL
Record frequency for automatic operation with interval
Printout
Printout type related to a working mode. Printout is activated by pressing PRINT key on
the overlay.
Options:
•
STANDARD PRINTOUT
Enables declaring printout content: HEADER, WEIGHING DATA and FOOTER.
Select <YES> parameter for the components that are to be printed on pressing
printout activating function key.
•
NON-STANDARD PRINTOUT
Enables selecting one of database-stored non-standard printouts form a list of
<PRINTOUTS> menu, or designing a unique printout which is automatically added
to the database.