- 91 -
Procedure:
1. Press grey workspace.
2. Settings, Keys, Information, Printouts, Series and Profiles submenus are displayed.
3. Press <Settings> key.
4. The display indicates functions related to differential weighing process.
Differential weighing process functions:
Threshold:
mass value (maximum mass of a sample, e.g. filter)
Ambient conditions:
use this function to force operator to enter temperature and humidity prior to
weighing operation: <Weighing A, Weighing T+A, Taring T>
•
NO – value not required;
•
ONLINE – values taken form ambient conditions module cooperating with the balance
•
VALUE – values taken form device other than ambient conditions module must be entered
manually.
Maximum number of weighing records:
use this function to declare how many times final weighing
<B> is to be repeated – 5 times maximum, the parameter setting is valid for all measuring series.
For overview of remaining functions go to section 15.7. “Additional Parameters for a Weighing
Process”.
25.2. QUICK ACCESS KEYS
Each working mode features automatically displayed set of default keys. The set can be modified by
assigning out-of-list quick access keys to on-screen keys. Such operation requires particular
permissions level.
This section presents only those buttons that are characteristic for Differential Weighing mode.
Weighing A
Press to start weighing initial mass <A>. The process is carried out as a separate
operation.
Weighing (T+A)
Press to start sample container taring which is next followed automatically by sample
weighing (non-separable operations). On initiation you are asked to set sample’s name.
Tare (T)
Press to start sample container taring. The process is carried out as a separate
operation. On initiation you are asked to set sample’s name.
Weighing B
Press to start weighing final mass <B>. The button triggers differential weighing mode.
Series
Press to select a sample for which the differential weighing process is to be carried out.
Sample
Press to select a sample for currently carried out process of a particular series.
Copy tare
Press to copy tare value, determined for a given sample, to all current series samples
(for which the tare value has not been assigned yet).
Delete value
Press to delete the most recently recorded mass value (tare, weighing A or weighing B).
Add sample
Press to add new sample, window with an on-screen keyboard is displayed, use it to
provide new sample name.
In order to add a new sample first it is necessary to select series for which the sample is
to be added. Mind not to start weighing operation before selecting the series and adding
the sample.