- 87 -
List of pipette parameters:
•
Name: press “Name” field, a window for entering pipette’s name opens.
•
Code: enter pipette’s code;
•
Model: enter pipette’s type;
•
Tip: enter tip’s name;
•
Type of volume: ADJUSTABLE / FIXED;
•
Nominal volume: nominal volume value;
•
Minimum volume: minimum volume value, for fixed volume pipette enter <0>;
•
Number of channels: specify number of pipette’s channels, for a single channel pipette enter
<1>;
•
Type: NONE/A/D1/D2. Pipette type adheres to standard guidelines. Since error values depend
on pipette’s type it is necessary to set the type if calibration process is to comply with ISO 8655
standard.
•
Tested volumes: press “Tested volumes” field, a window with a list of calibrated volumes opens
(in case of a new pipette, the list is empty), add volume to be tested, and define error value for
each tested volume:
Procedure:
•
Press <
Add > button.
•
A window with an on-screen numeric keyboard is displayed.
•
Enter tested volume in [µl] and press <
> button for confirmation.
•
A new record is automatically added, the record features recommended error values.
In order to change error values press a field with pipette volume (each field can be edited,
you can enter error values of your choice).
Caution:
While in calibration process, the order of tested volumes corresponds to the order in which they
are entered.
•
On setting correct values return to the main window.
24.4. PRINTOUTS
Printouts option has been designed to enable you to determine content of particular standard
printouts and to define non-standard printout.
Standard Printout
Standard printout comprises four internal blocks, each
featuring different variables. Each variable can be attributed
with parameter YES – i.e. printed, or NO – i.e. not printed.
Settings for header printout, weighing printout, and footer
printout are specified in section 14.5; settings for Pipettes
Calibration report are presented below.
Report content:
•
Working mode
•
User
•
Customer
•
Temperature
•
Humidity
•
Pressure