- 93 -
Upon selecting a series, press one of the following buttons determining how the differential
weighing is to be carried out:
Weighing A
Press to start weighing initial mass.
Weighing (T+A)
Press to start sample container taring which is next followed automatically by sample
weighing (non-separable operations).
Tare (T)
Press to start sample container taring.
Weighing B
Press to start weighing final mass. The button is enabled only for a series containing
samples for which weighing A has been carried out.
Weighing A
•
On pressing Weighing A button, the balance
searches samples list in order to find the first
sample for which weighing <A> is still to be
carried out. If there are no such samples the
balance displays a message informing you that
the operation is disabled.
•
If the operation is enabled then new data on
initiated process is displayed in the grey
workspace. You are guided through the process
by means of prompts visible either in the
workspace or bottom bar.
•
Load the weighing pan with a sample and press
<
> button.
For <Ambient condition>
parameter set to <Value> parameter a window for
entering temperature and humidity values is
displayed. For <Ambient condition> parameter set
to <ONLINE> parameter
a window for
confirmation of values taken from ambient
conditions module is displayed.