- 83 -
MYA 4Y balance with a set for pipettes calibration:
Installation procedure:
1. Dismantle the weighing pan and the anti-draft chamber
2. Place the following elements inside:
•
anti-anti-draft chamber (1)
•
bottom ring (2)
•
weighing pan (3)
•
glass vessel (5)
•
evaporation ring (6)
•
glass cover (7)
•
supplementary cover (8)
•
or glass cover (4)
XA 4Y balance with a set for pipettes calibration:
Installation procedure:
1. dismantle the weighing pan and the anti-draft chamber,
2. place centering ring (8) inside the chamber,
3. put glass ring (7) onto the centering ring,
4. place weighing pan (6) inside the glass ring,
5. put top ring (5) onto the glass ring,
6. put evaporation ring (4) onto the top ring (5)
7. insert calibration vessel (3) into the evaporation ring,
8. put the glass cover (2) onto the top ring,
9. put glass supplementary cover (1) onto the glass cover.
Balance with pipette calibration set installed.