- 132 -
Distribution probability chart:
30.14. MINIMUM SAMPLE WEIGHT
Minimum sample weight database contains data on declared methods and minimum sample weights.
Caution:
Only authorized RADWAG personnel can enter new minimum sample weight records or make changes in
those stored in the database.
Procedure
1. Enter submenu <
Databases> and press <
Minimum sample weight>.
2. Press <
Add > button to add new minimum sample weight.
3. If a record on minimum sample weight is already stored in the database, then press field
with its name to enter its details that can be modified.
List of parameters determined for a minimum sample weight:
Name
– name of method used for determining minimum sample
weight.
Code
– method code.
Description
– method description
Subsequent control
– a field for entering expiry date of the minimum sample
weight. 2 weeks before the expiry date, an icon of a clock
displays next to the status icon. Contact RADWAG
representative to carry out necessary changes in minimum