- 53 -
Radwag has designed analysis method and selected permissible vibrations thresholds
respectively to make signalling device operation both effective and convenient (not
troublesome). Balance screen displays balance-registered vibrations level in percent.
Vibrations threshold cannot be changed, however the operator can modify settings in a way
ensuring that alerts are displayed in the most convenient way, depending on the needs. By
default the set value is 100%.
Example of settings modification:
For better understanding, analyse the following example (the values used below are
exemplary ones, they were not taken form the balance).
1.
Assumed vibrations threshold set by default on 4Y series balance: 100mm/s
2
.
2.
Max vibrations value: 100% (alert display threshold – pictogram colour change, from
green to red). Pictogram colour is green for vibrations within 0 - 100mm/s
2
range, for
vibrations value of 101mm/s
2
, the pictogram changes to red.
3.
Max vibrations value: 150%. Pictogram colour is green for vibrations within 0 - 150mm/s
2
range, for vibrations value of 151mm/s
2
, the pictogram changes to red.
Vibrations are signalled by a respective pictogram displayed in the weighing result window.