- 173 -
2 – fast+reliable
3 – reliable
CAUTION!
The numbering is assigned to a particular value release option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, value
release settings are assigned to the working mode
An example:
Command:
ARS_2 CR LF
– set value release parameter to fast+reliable
option
Response:
ARS_OK CR LF
– fast+reliable option set
Set last digit
Format:
LDS_n CR LF
Response options:
LDS_OK CR LF
– command carried out
LDS_E CR LF
– error in-course of command execution, no parameter or
incorrect format
LDS_I CR LF
– command understood but not accessible at this moment
n
– parameter, decimal value determining last digit settings
n →
1 – always
2 – never
3 – when stable
CAUTION!
The numbering is assigned to a particular last digit option and it is identical for all balance types.
The command changes settings for a current working mode if, for a particular balance type, last digit
settings are assigned to the working mode.
An example:
Command:
LDS_1 CR LF
– set last digit option to value ‘always’
Response:
LDS_OK CR LF
– ‘always’ value set
Give accessible units
Command overview: Command returns units available for a particular device and for a current
working mode.
Format:
UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK CR LF
– command carried out, returns accessible units
UI_I CR LF
– command understood but not accessible at this
moment
x
– unit symbols, separated by means of comas
x
→ g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI CR LF
– return available units
Response:
UI_”g, mg, ct”_OK CR LF
– response: available units
Set current unit
Command overview: Command sets current unit for a particular device.
Format:
US_x CR LF
Response options:
US_ x_OK <CR><LF
– command carried out, response: current unit
US_E CR LF
–error in-course of command execution, no parameter or
incorrect format
US_I CR LF
– command understood but not accessible at this moment
x
– parameter, units symbols: g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola,
msg, u1, u2, next
CAUTION! if x=next, the command swaps to another avaialable unit on the list (it simulates „
”
button pressing).
An example:
Command:
US_mg CR LF
– set „mg” unit
Response:
US_mg_OK CR LF
– „mg” set as a current unit