- 66 -
19.3. BARGRAPH - A TOOL FOR FUNCTION INTERPRETATION
The DOSING and CHECKWEIGHING functions are aided by a graphic indication, i.e. bar graph.
Example of simultaneous operation of DOSING and CHECKWEIGHING functions.
•
Checkweighing thresholds Low Limit = 90%, High Limit = 110%.
•
Target value = 105% ; tolerance = 5%.
•
Reference mass = 19.986 g <
>
19.4. SUPPLEMENTARY SETTINGS FOR PERCENT WEIGHING MODE
The supplementary settings enable adjusting the working mode to your needs and requirements. For
information on how to access the settings go to section 15.7.
20. DENSITY
<
Density>
is a working mode enabling to determine density of:
•
solid bodies,
•
liquids,
•
air (available for XA 4Y),
•
various substances with use of pycnometer.
In order to operate this mode, an optional density kit (supplementary equipment), appropriate to
a model of used balance, is required. Prior to installing the kit it is necessary to remove the weighing
pan and an anti-draft chamber.
Density mode screen.
1. Setup – press to access balance menu
2. Print header – press to print data
declared in the header
3. Print footer – press to print data
declared in the footer
4. Density of solid
5. Density of liquid
6. Density of air
7. Pycnometer
Summary of Contents for UYA 4Y
Page 8: ...8...
Page 9: ...9 1 GENERAL INFORMATION 1 1 DIMENSIONS Ultra microbalances and microbalances...
Page 10: ...10 MYB 4Y Microbalances Indicator XA 4Y series balances...
Page 11: ...11 XA 4Y F series balances...
Page 195: ...195...