1601 MP 8, 1602 MP 8 und 1612 MP 8 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Installation and Operating Instructions 

 
 
 
 
 
 
 
 
 
 

Summary of Contents for 1601 MP 8

Page 1: ...Electronic Analytical Balances 1601 MP 8 1602 MP 8 und 1612 MP 8 Installation and Operating Instructions ...

Page 2: ...2 Electronic Analytical Balances 1601 MP 8 1602 MP 8 and 1612 MP 8 ...

Page 3: ...rating temperature 10 o C to 40 o C Sensitivity drift in the 10 o C 30 o C range 2 10 6 K Deviation from result when tilted 1 1000 mg 0 1 Calibration weight extern 100 g see accessories built in 150 g weight standard feature tol 0 1 mg built in 30 g 100 g standard feature tol 0 002 0 2 mg Data output optional Below balance weighing device standard feature Pan diameter mm 90 clearance above pan mm ...

Page 4: ...voltage Pull out and turn the voltage selector 16 and plug it in such that the desired voltage is aligned with the arrow mark Be sure to write the new setting on the manufacturer s label 17 Line voltage selector 20 Compare factory set voltage to avaible local line voltage and reset if necessary Installing the pan Proper sequence shield plate 14 pan 10 protective ring 13 ...

Page 5: ...ON all components subject to wear are switched off stand by operation after the autocheck balance is instantly ready to operate Leveling Level the balance using leveling screws 5 and level indicator 9 On model 1612 MP8 the level indicator is to right of the pan 2 Rang switching on model 1612 MP8 only macro range 160 g 0 1 mg in pos 160 g ...

Page 6: ...ernal calibration weight on model 1602 MP8 150 g and model 1612 MP 8 30 100 g 3a Position CAL application of internal calibration weight on 1612 MP8 CAL can only be reached via switch position 30 g 4 Position TA Before transporting 1612 MP8 make sure mode switch is set to TA 3 ...

Page 7: ...ll power plug Autocheck checks the balance for proper function Defect message appears Taring Press tare button 6 display shows 0 0000 g as code number in display 8 It tells the service technician where to troubleshoot Stability control 7 When stability symbol g appears readout is reliable ...

Page 8: ...bration weight only see accessories 100g 0 15mg Remove cap from calibration button 4 clear pan tare 6 and wait for stability 7 Press calibration button 4 Display shows C and the busy symbol 21 if display shows CE repeat calibration starting from tare Sensitivity adjustment 1602MP8 using the built in calibration weight Tare 6 At stability 7 set operating mode switch 3 to CAL Display shows C and the...

Page 9: ...0 1 mg Clear pan Tare 6 Wait for stability 7 Slightly depress operating mode switch 3 and carefully set it to Cal 30 g see picture 7d Display shows C and the busy symbol 21 lf you get CE display restart calibration from tare 6 Place 30 g calibration weight tol 0 02 mg on pan When CC is displayed balance will automatically return to weighing function 2 Macro range 160 g 0 1 mg Tare 6 and wait for s...

Page 10: ...For Optimum adaption of the balance to your specific application and operating environment a menu program provides selection of the ideal combination from the following parameters 1 digital filter 2 stability range 3 display format 4 tare condition 5 Auto zero Additional Parameters for data Output format and calculation programs with external keyboards are available Please refer to accessories lis...

Page 11: ...gits C 1 2 4 2 digits C 1 2 5 4 digits C 1 2 6 8 digits C 1 2 7 16 digits C 1 2 8 32 digits C 1 2 9 64 digits Code Display format C 1 3 1 last decimal ON C 1 3 2 last decimal OFF C 1 3 3 last decimal at stability C 1 3 4 all decimal at stability Code Tare mode C 1 4 1 without stability C 1 4 2 at stability Code Auto zero C 1 5 1 ON C 1 5 2 OFF Code C 4 1 1 Program lock OFF C 4 1 2 Program lock ON ...

Page 12: ...rmination 6080 External calibration weight for 1601 MP8 707219 Data output 735736 External tare button connects to data output 7226 Printer Data Print 7279 External Keyboard Data Input 73432xxx Combination Data Input Data Print 73802xxx Converter from RS 232 V 24 S to JEC 725316 BCD parallel 725318 RS 422 725319 analog 7287 Programmset Programme auf Anfrage Loading through top deck Lift up the but...

Page 13: ......

Page 14: ... Copyright by Sartorius AG Göttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartorius AG Sartorius AG reserves the right to make change to the technology features specification and design of the equipment without notice ...