QIAcuity User Manual 06/2022
30
Removing the protective film from the QIAcuity touchscreen
Carefully peel off the protective film from the QIAcuity touchscreen.
Removing the shipping fixation screw
Access the back of the instrument and remove the shipping fixation screw using a 4 mm hex wrench.
Store the fixation screw on a safe place in case it is needed at a later point in time. The hole in the
housing for the fixation screw shall be closed with the dust cap that is provided with the accessories
of the instrument (cat. no. 9026772).
Back of QIAcuity.
Connecting the power cord to the back of the QIAcuity
1.
Remove the power cord from the foam packing material on top of the QIAcuity.
Note
: Only use the power cord that is supplied with the QIAcuity.
2.
Check that the voltage rating on the label found at the back of the QIAcuity matches the
voltage available at the installation site.
3.
Connect the power cord to the power inlet of the instrument and connect the cable to the wall
power outlet.
4.
Turn on the power switch at the back of the instrument.
Summary of Contents for QIAcuity 911000
Page 2: ...QIAcuity User Manual 06 2022 2...
Page 68: ...QIAcuity User Manual 06 2022 68...
Page 240: ...QIAcuity User Manual 06 2022 240...