WARNING
Explosive atmosphere
The QIAcuity is not designed for use in an explosive atmosphere.
[W21]
CAUTION
Risk of overheating
To ensure proper ventilation, maintain a minimum clearance of
10 cm at the sides and rear of the QIAcuity.
Slits and openings that ensure the ventilation of the QIAcuity must
not be covered.
[C9]
WARNING
Risk of personal injury and material damage
The QIAcuity is too heavy to be lifted by one person. To avoid
personal injury or damage to the instrument, do not lift the
instrument alone. The bottom plane shall be used for lifting. Do not
lift at the touchscreen.
[W2]
CAUTION
Damage to the instrument
Direct sunlight may bleach parts of the instrument and cause
damage to plastic parts.
The QIAcuity must be located out of direct sunlight.
[C6]
Power requirements
The QIAcuity operates at 100–240 V AC, 50/60 Hz, 1500 VA (max.)
Ensure that the voltage rating of the QIAcuity is compatible with the AC voltage available at the
installation site. Main supply voltage fluctuations should not exceed 10% of nominal supply
voltages.
WARNING
Damage to electronics
Before powering ON the instrument, make sure that the correct
supply voltage is used.
Use of incorrect supply voltage may damage the electronics.
To check the recommended supply voltage, refer to the
specifications indicated in the type plate of the instrument.
[W9]