QIAcuity User Manual 06/2022
283
System status environment on the QIAcuity Eight after an error occurs.
4.
To clear an error, tap
Clear error
.
5.
If the error that occurred affects the tray(s), tap
More details
. To clear a tray-related error, tap
Clear error in the dialog. The dialog box contains five items that can be cleared for each tray,
such as motor and slot numbers (based on Insrument Version).
Note
: In QIAcuity Eight, the
More details
button is located in the Upper Tray and Lower Tray
panes. In the QIAcuity Four and QIAcuity One, the
More details
button is located in the Tray
pane.
6.
Restart the instrument. The instrument initializes and all modules are returned to their home
positions.
Note
: If the affected module is not working after you cleared the error and restarted the
instrument, contact QIAGEN Technical Services.