GB 03379 - Edition 02- June 09
1/1
D 200
Diagnosis and troubleshooting
I
NCIDENT
C
AUSE
R
EMEDY
No event occurs after
power on
No mains current
Check that the pump is
powered by a 48 V DC power
supply (the blue lit lights on).
Check the remote control
configuration of the pump
(
C800
: SEL10).
Contact the Customer Service.
Isolate the product from the
various energy sources before
any maintenance operations.
During the speed rise, the
pump starts to vibrate
Abnormal working
Check that the pump is rigidly
fixed to the frame;
Check that there are no
vibrations on the pump frame;
Check that the attachment
screws are correctly fastened
(problem can be caused by an
anti-vibration flagstone);
Contact the Customer Service.
The red light (on the front
panel) lights on
The pump is faulty
Connect the serial link;
Enter the command: #DEF
(
C800
);
Example of faults
- unbalance fault
- magnetic bearing
- displacement fault
- over temperature
Example of warnings
- int. comm
- power volt
- OBC temp.
Enter the command : #LEV10
(
C800
)
Switch off then switch on the
power supply;
If the fault persists contact the
customer service.