2/6
C 800
Detailed description
of RS 232 and RS 485 commands
GB 03376 - Edition 03 - June 2012
ECH
Enables or disables command echoing
Syntax
#adr
ECH
ON<CR>
enables all characters received to be echoed over the serial port
(RS 232 only).
or
#adr
ECH
OFF<CR>
disables all characters received from being echoed over the serial port.
Result
#adr,ok
Comment:
- This command is disabled in RS 485 operation, the value OFF is
required.
IDN
Identifies the device which is communicating, and its software
version
Syntax
#adr
IDN
<CR>
Result
#adr, ATH500M - Vx.yy.zz
Returns the type and the software revision of variable drive superviser.
LEV10
Returns the state of the parameters defined by SET
Syntax
#adr
LEV10
<CR>
Result
#adr,nnnnn,sssss,00000,0,ccccc,eeeee,00000,0000,0000,jj,kk,lll,mmm
Returns current values:
nnnnn
: nominal speed set point (in rpm)
sssss
: stand-by speed set point (in rpm)
ccccc
: pump working time (in hours)
eeeee
: electronic working time (in hours)
jj
: speed threshold for relay (3 to 50%)
kk
: temperature set point (30 to 65%)
lll
: bearing threshold (0 to 100%)
mmm
: bearing current value (0 to 100%)
Note: 0 = not used
NSP
Switches the speed set point to the nominal speed value
Syntax
#adr
NSP
<CR>
Result
#adr,OK
The speed set point for the pump is set to its nominal value.
adixen Vacuum Products - Operating instructions - ATH 500 M-MT