GB 03367 - Edition 03 - Jan. 12
1/5
B 300
Pump connections to an installation
Equipment
installation
conditions
Inlet
Inlet
Inlet
Inlet
The pump can operate in
any position
When the pump is connected according to the manufacturer
recommendations (
B300) , in case of overpressure resulting from an
incident, the pump enclosure can withstand a maximum static pressure of
11 bar for 1 minute.
Pump installation :
- connect the pump to the customer’s handling device using the threaded
holes located at the pump housing (
A400).
The equipment frame on which the pump is installed must be sufficiently
rigid to absorb the kinetic energy of the rotor in case of pump rotor crash.
For this, take into account :
• the maximum loads to calculate the equipment attachment devices,
• the flange dimensions,
• the quality and the number of bolts.
Do not use bellows on the pump inlet flange and the chamber Reducing
adapter must be correctly dimensioned.