C 100
Safety instructions for product use
GB 03374 - Edition 03 - Jan 12
Standard precautions before any maintenance operations:
Before performing a maintenance operation, stop the pump. When the
pump is at rest, switch off the pump by setting the controller main switch
to «0», wait 5 minutes before disconnecting the main cable.
If this last one remains connected, some components will still be energized.
Be sure that the controller status is visible from the operator otherwise
disconnect the cable from the pump.
Located on the controller, this label indicates that the controller musn’t be
disconnected when the pump is running.
LOCK
- Use only specific
cable.
- Lock connector
before use.
- Do not unplug when
the pump is on.