GB 03376 - Edition 03 - June 2012
3/6
C 800
Detailed description
of RS 232 and RS 485 commands
OPT
Used to select possible user choices
Syntax
#adr
OPT
xx,n<CR>
Result
#adr,OK
xx = OPTION OF
PARAMETERS
n = value
14: Remote mode
n = 0 : Keyboard (for HHR)
n = 1 : remote hard (give control to remote)
n = 2 : serial link (give control to RS 232/485)
n = 5 : Profibus
25:
Inhibition
mode
n = 0 : inactive
n = 1 : active inhibition
(the controller doesn’t supply the motor)
33 : Contact mode
n= 0 : ROTATING
(closed contact if speed>120rpm
n= 1 : ACCELERATING
(closed contact in phase of progress)
RDI
R
eturns the serial number of the product
Syntax
#adr
RDI
<CR>
Result
#adr,A123456789<CR>
RPM
Defines the speed set point in stand-by value
Syntax
#adr
RPM,
nnnnn<CR>
Result
#adr,OK
or
#adr,ErrX
X = 1: out of range
X = 2: parameters
X = 3: context (not in standby mode)
SBY
Switches the speed set point to the stand-by value
Syntax
#adr
SBY
<CR>
Result
#adr,OK
Resets the stand-by speed to its Standby stored value, and allows it to be
modified if a
«RPM»
command is sent.
This configuration is automatically stored in user memory.
adixen Vacuum Products - Operating instructions - ATH 500 M-MT