GB 03371 - Edition 04 - Jan. 12
1/2
B 430
“Remote control” connector wiring
When units containing control circuits are equipped with dry contact outputs,
it is the responsability of the customer to use these outputs in compliance
with extra low voltage installation and security standards. It concerns
Remote, Profibus and RS connectors, excepted main power connector.
Provided power supply by
the pump:
10
5
(+)
(-)
12 V DC - 20 mA
GND
15
5
10
15
Input
Return
Start/Stop
1
6
11
Standby
12 V DC - 20 mA
GND
REMOTE
The inputs dry contacts Sub D 15 Pts Fem:
External power supply:
(+)
(-)
12/24 V DC
15
Remote function if
“Remote hard“ mode selected
Remote
start/stop
14(+) 15(-)
- Active signal starts the pump rotation.
- Inactive signal stops the pump rotation.
Remote
standby
13(+) 15(-)
- When the “Standby“ is validated, the rotational
speed corresponds to the selected speed.
- When the “Standby“ is disabled, the rotational
speed corresponds to the nominal speed.
The control by
voltage
12 V or 24 V DC
Customer wiring
Customer
wiring
Pump wiring
Pump wiring
adixen Vacuum Products - Operating instructions - ATH 500 M-MT
adixen Vacuum Products - Operating instructions - ATH 500 M-MT