GB 04243 - Edition 03- June 2012
1/1
C 300
Configuring the ATH 500 M-MT pump
for the application
Pump parameter configuration
PARAMETERS
RS COMMANDS
VALUES
FACTORY
CONFIGURATION
Modify the Standby speed
RPM and SBY
15 000 to 50 000 rpm
15 000
Modify the speed contact
threshold
SET 30
- 3 to -50%
- 3%
Modify the thermostating
threshold (only MT model)
SET 31
30 to 65 °C
65 °C*
Modify the bearing alert
threshold
SET 32
0 to 100%
20%
Remote mode
OPT 14
0 : Keyboard (HHR)
1 : Remote
2 : RS 232
5 : Profibus **
1 for standard pump
or
5 for Profibus
Inhibition mode
OPT 25
0 : inactive
1 : active
0
Thermostatage mode
OPT 29
0 : inactive
1 : active
0 : for ATH 500-M
1
:
for ATH 500-MT
Contact mode
OPT 33
0 : Rotating
1 : Accelerating
0
PARAMETERS
RS COMMANDS
VALUES
FACTORY
CONFIGURATION
Transmission speed
-
-
9600 bauds
Data length
-
-
8 bits
Parity
-
-
None
Number of STOP bits
-
-
1 bit
Data separating characters
SEP
0 to 255
44 (comma)
Number of controller in a
link
ADR
0 to 255
0
Authorize transmission on
STA at pre-set intervals on
the serial link, when ON is set
DLR
ON or OFF
OFF
Set interval transmission
DLI
0 to 4 min 15s
or 255 s
0 mn 1s
RS232/RS485 serial link settings
* Depends on the pump model
** Only with Profibus option
adixen Vacuum Products - Operating instructions - ATH 500 M-MT