GB 03381 - Edition 03 - Sept. 11
1/3
E 100
Shipping procedure
for contaminated pumps
Pumps to be shipped must initially be decontaminated then
pressurized with dry nitrogen (see procedure sheet 3/3).
To achieve this the user must have the following connection
accessories:
Inlet port
A
Don’t forget to fill in the «safety questionnaire» and return it to repair
service center (see model of document at the end of the manual).
Study the safety instructions related to preventive maintenance
D 100.
Closing kit
DN160 ISO F
DN160 ISO K
DN100 ISO K
P/N
114501
114502
118377
Closing kits including o-ring, screws and nuts inlet blank flange and
hoisting rings can be supplied upon order. For other inlet part types
contact the customer service.
Whenever you return the product to an adixen repair service center, you
must comply with the requirements of the «Procedure for returning adixen
products», fill in the safety questionnaire available in appendix
G200 of the user’s manual.
Risk of injury by cutting.
Contact with the pump rotor cell may cause cuts. Alternatively, protective
gloves may be worn when servicing the product.
The user must stick this label on the product to warn against
pumped process gas that could be dangerous and toxic and could
cause severe injuries or death. It precises that any preventive
maintenance operation can only be performed by trained
personnel.
FLAMMABLE, CORROSIVE AND TOXIC
CHEMICALS LOCATED WHITHIN THE ENCLOSURE
Explosure may result in severe injury
or death.
Preventive maintenance must be done
by trained personnel only.
WARNING
adixen Vacuum Products - Operating instructions - ATH 500 M-MT
adixen Vacuum Products - Operating instructions - ATH 500 M-MT