4/6
C 800
Detailed description
of RS 232 and RS 485 commands
GB 03376 - Edition 03 - June 2012
SEL10
Returns the state of the parameters defined by OPT
Syntax
#adr
SEL10
<CR>
Result
#adr,0,0,1,0,r
not used
r: Returns remote mode choice
r=0:
keyboard
r=1: remote hard
r=2: serial link
r=5:
Profibus
Result
#adr,OK
Note: 0 = not used
1 = not used
SEL 20
Returns the state of the parameters defined by OPT which are
not defined in SEL 10
Syntax
#adr
SEL20
<CR>
Result
#000,a,b,c,d,e,f,g,h,i,j,k,l,m,n,o,p,q,r,s,t,u,v,w,x,y,00,11,22,33,44,55<CR>
f :opt 25
j :opt 29
n :opt 33
SEP
Defines the character which separates the parameters in a
reply
Syntax
#adr
SEP,
nnn<CR>
nnn: 3-digit decimal value of the ascii code of the desired charac-
ter (with leading zeros).
condition: 000
≤
nnn
≤
255
Result
#adr,OK
or
#adr,ErrX if error
Allows the user to select the character which separates the parameters retur-
ned by the
DLR
STA
and
LEV
commands.
Default value: comma «,» ascii code = 044
If ok, the selected value is automatically stored in user memory.
SET
Defines the internal operating parameters
Syntax
#adr
SET
10,ccccc<CR>: utilization time of the pump (in hours)
#adr
SET
30,jj<CR>: speed threshold for relay (3 to 50%)
#adr
SET
31,kk<CR>: thermostating threshold (30 to 65 °C)
#adr
SET
32,lll<CR>: bearing threshold (0 to 100%)
#adr
SET
33,MMM<CR>: bearing current value (0 to 100%)
Result
#adr,OK or ErrX
adixen Vacuum Products - Operating instructions - ATH 500 M-MT