2/4
A 400
Technical characteristics of the pumps
GB 03362 - Edition 03 - Jan 12
Characteristics
UNITS
ATH 500 M
ATH500MT
Run-up TIME
mn
< 2.0
Power consumption max. starting up power
watt
560
Nominal power
watt
100
Standby power
watt
< 50
Controller power supply
48 VDC
Maximum leakage current
mA
not applicable ; polarity connected to the
earth
Recommanded cooling water flow rate
l/hour
60
Maximum water line pressure
bar
7
Cooling water temperature
(5)
°C
15 - 25
OBC (on-board-control) control option
Profibus
in option
no
Remote
yes
Devicenet
no
Weight
kg (lb)
17 (37)
18(39)
Recommended backing pump
ACP 40
Air cooling
option
yes
no
Environmental conditions :
Use of the product
Ambient operating temperature
Maximum altitude
Transient overvoltages typically present on
the mains supply
Pollution degree applicable
Maximum relative humidity
Accept temporary overvoltages on the
power line
•
•
•
•
•
•
•
-
°C
m/ft
-
-
-
Indoor only
5 < T < 45
2000/6562
Category II
2
Maximum relative humidity of 80 % for tem-
perature until 31°C with a linear decrease
until 50 % relative humidity at 40°C
Three phase motor characteristics
(Max. values) :
Voltage between phases
Supply frequency
Phase current
•
•
•
Vrms
Hz
A
29
1666
9.5
The performances
of the pumps
(5) > 25 °C : apply a derating. Change the operating conditions to reduce the nominal power. Contact the Customer Service.
Value for pump and controller. Not all the leakage current is necessary evacuated through the earth wire; it depends on
the pump earth connection regarding the equipment to which the pump is connected.
adixen Vacuum Products - Operating instructions - ATH 500 M-MT
adixen Vacuum Products - Operating instructions - ATH 500 M-MT