2/4
G 1000
Alcatel Vacuum Technology France - ATH 200 User’s Manual
GB 04070- Edition 03 - Jan. 12
D I S P L AY
S ET U P
S E R N U M
P R E V.
-
+
E N T E R
- Connect the hand held remote to the controller on «
Service connector
».
- Supply the pump controller.
Enter the sub-menus by pressing :
Access to the menus
Before using the HHR, check that the electrical connections have been made
between the pump and the controller (refer to B400).
HHR Display
initialization
The initialization time is
approximately 15 seconds.
The controller performs a self-test
and identifies the pumps to which
it is connected.
The equipment is identified, the
program version is displayed, than
communication test is done,
In the meantine, the Indicator lights
are tested by lighting in succession.
Finally, the working screen
corresponding to the connected
pump is displayed.
HHR V01.00.08
Checking procedure
Display and/or select
the parameters to be
monitored
(see page 3)
Access the parameter
programming
(see page 4).
Access the hardwire component
serial number
(see page 4).
Hand held remote start up
HHR V01.00.08
Collecting data
KEYB 0H
CTRL VXX.XX.XX
ATH 1234 M
... ....Ready to start..........
adixen Vacuum Products - Operating instructions - ATH 500 M-MT