GB 04734 - Edition 01 - Jan 12
1/1
B 333
Assembly of the purge kit
1
2
3
4
6'
5
6
7
N2
Installation purge kit
on ATH 500M
1 - Replace the screws M4-16 of electronic by the screws 5 and the
washers 4 and 7 .
2 - Remove the silencer or purge plug from the purge port.
3 - Screw the union 6’ on the purge port.
4 - Install the assembly 2 on the top of the electronic box using
screws 5 through the plate.
5 - Screw N2 gas line in the union 6’ .
6 - Tighten the nuts 3 with the washer 4 .
7 - Test the vacuum tightness.
1/8 BSPP connection
Integrated electronic
adixen Vacuum Products - Operating instructions - ATH 500 M-MT
adixen Vacuum Products - Operating instructions - ATH 500 M-MT