Unbalanced force
rejection control
The
unbalanced force rejection control
is an electronic
function, that monitors the rotor position, allowing it to rotate
in its own axis of inertia.
Changes in the rotor balance, due to deposit built-up during
the life time of the pump, are automatically compensated by
the
unbalanced force rejection control.
It ensures the lowest possible levels of noise and vibration.
Stator geometric axis
Rotor geometric axis
Rotor inertia axis
A 200
Control loop of the pump
2/2
GB 02721 - Edition 04 - Aug 07
adixen Vacuum Products - Operating instructions - ATH 500 M-MT
adixen Vacuum Products - Operating instructions - ATH 500 M-MT