6/6
C 800
Detailed description
of RS 232 and RS 485 commands
GB 03376 - Edition 03 - June 2012
STA
(continued)
Returns the status of the internal dynamic parameters
(continued)
g
0 = OK
1 = ALERT
2 = FAULT
0
0
1
0
2
0
3
0
4
0
5
0
over-current/sensor
6
0
7
0
mag. suspens.
8
0
Power voltage
9
0
10
0
Yh radial
11
0
Xh radial
12
0
Yb radial
13
0
Xb radial
14
0
Z axial
15
0
bearing
Bearing change
16
0
electronic temp.
Electronic temp.
17
0
pump temp.
Pump temp.
18
0
19
0
20
0
Self check
21
0
22
0
23
0
24
0
TMP
Defines the operating state of the turbomolecular pump
Syntax
#adr
TMP
ON<CR>: start pump rotation
#adr
TMP
OFF<CR>: stop pump
Result
#adr,OK
or
#adr,Err3 if the pump is already in the state requested
(context error)
VER
Defines the version of electronics «extension» + «frequency
converter» + «magnetic bearing»
Syntax
#adr
VER
<CR>
Result
#adr,Interface : VX.YY.ZZ, cartridge :VX.YY, Type XXXXX<CR>
Interface VX.YY.ZZ:
interface board software version (Front panel)
Cartridge : VX.YY :
interface board software version
Type:xxxxx
type of connected pump (code)
adixen Vacuum Products - Operating instructions - ATH 500 M-MT