Set up the Splitless Mode for a CAP or PSS Injector . 99
At the end of a run, the split flow resets to the original setting. Setting a Value of 50 mL/min
minimizes contamination by keeping the splitter on during cool-down.
NOTE: If you wish to conserve carrier gas, set the initial flow to 0.
CAUTION
ATTENTION
If you are using pressure pulse injection, you must set the split flow on
just prior to when the inlet pressure begins to drop to the operating
setpoint. For example, if the inlet pressure is programmed to start
decreasing one minute after injection, use a split flow event time of 0.98
minutes.
Si vous utilisez une injection d'impulsion de pression, vous devez régler le
flux divisé juste avant que la pression d'entrée commence à tomber à la
consigne de fonctionnement. Par exemple, si la pression d'entrée est
programmée pour commencer à diminuer une minute après l'injection,
utilisez un temps d'événement fractionné de 0,98 minute.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...