NPD Overview . 159
Activating the NPD Bead
The bead requires an activation step prior to use.
Do not inject a sample during Activation.
To Activate the bead, carefully follow this procedure in sequential steps below:
1.
Turn the bead potentiometer dial clockwise to apply current to the bead.
2.
Slowly increase the NPD control voltage potentiometer to a setting of 700.
NOTE: The bead will gradually acquire an orange glow at this point.
3.
Wait for 5 minutes then increase the potentiometer to 750.
4.
After 1 minute move the potentiometer to 800.
5.
After 1 minute move the potentiometer to 850.
6.
After 1 minute move the potentiometer to 900.
7.
After 1 minute reduce the potentiometer to 850.
8.
After 1 minute reduce the potentiometer to 800.
9.
After 1 minute reduce the potentiometer to 750.
10.
After 1 minute reduce the potentiometer down to 700
11.
For Capillary columns:
Wait 10 minutes, go to the NPD detector screen, then adjust the
potentiometer so that the NPD baseline response reads around 0.5 mV on Range 1. Then set the
desired column flow for the application to be run.
For Packed Columns:
Wait 10 minutes, go to the NPD detector screen, then adjust the
potentiometer down to 500 or below to drop the bead below the active level. Remove the Detector
Cap Nut (P/N N9300061) and reconnect the column to the detector. Set the desired column flow
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...