160 . Clarus 590 GC User’s Guide
for the application to be run, then again increase the bead current potentiometer to give a reading
around 0.5 mV on Range 1.
12.
Allow the baseline to stabilize for 30 minutes.
13.
Using the potentiometer, adjust the baseline between 0.4 and 0.6 mV at Range 1 and the bead
ready to use.
14.
After conditioning the bead remove the Swagelok plug from the detector fitting and reconnect the
column.
15.
Make sure the column flow is properly set. Touch the
signal
icon
to display the detector
background produced by carrier gas flow. The bead background will be less due to cooling effects.
16.
Turn the bead potentiometer dial clockwise until you achieve a reading of 0.25 mV above the "off"
reading with the detector range set to x1.
17.
Once the system has stabilized, it is ready for operation.
If you are unable to achieve adequate sensitivity for your standard, increase the background of the
NPD. Increasing intervals of 0.25 mV are recommended.
Adjust the background setting to achieve the sensitivity required for your samples. For example, if
you have a concentrated sample, such as drugs, it can be analyzed with a background of 0.25 –
0.50 mV. If you have a low-level sample, such as pesticides, it may require a setting as high as 1 –
2 mV. To get the maximum life from your NPD bead, operate at the lowest setting to achieve your
required sensitivity.
NOTE: (1) The bead will be stable after operating for several hours, but it will drift in time. At the start of each
day, adjust the background reading to the setting you are using and allow a few minutes for it to stabilize.
(2) Due to the loss of alkali metal with use, the nature of the bead is to drift over time. We strongly
recommend using an internal standard for quantitative analysis.
(3) The bead can operate at higher background settings. The higher the setting, the greater the signal and
noise. Therefore, the signal to noise ratio will not increase dramatically. Operate at the lowest possible
setting to achieve the required sensitivity; this will also prolong the life of the bead.
Summary of Contents for CLARUS 590 GC
Page 1: ...CLARUS 590 GC User s Guide GAS CHROMATOGRAPHY ...
Page 9: ...About This Manual 9 Flame Ionization Detector FID 270 TCD 273 Autosampler 277 Index 279 ...
Page 10: ...10 Clarus 590 GC User s Guide ...
Page 11: ...1 Introduction ...
Page 12: ...12 Clarus 590 GC User s Guide ...
Page 21: ...2 Touch Screen Navigation ...
Page 22: ...22 Clarus 590 GC User s Guide ...
Page 48: ...48 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 50: ...50 Clarus 590 GC User s Guide PPC Active Method Configuration Screen Manual Pneumatics ...
Page 59: ...3 Using the Active Method ...
Page 60: ...60 Clarus 590 GC User s Guide ...
Page 109: ...4 Setting Up the Detectors ...
Page 110: ...110 Clarus 590 GC User s Guide ...
Page 197: ...5 Using the Method Editor ...
Page 198: ...198 Clarus 590 GC User s Guide ...
Page 207: ...6 Using the Tools Menu ...
Page 208: ...208 Clarus 590 GC User s Guide ...
Page 224: ...224 Clarus 590 GC User s Guide ECD PID NPD FPD ...
Page 225: ...Detectors 225 Output Configured ...
Page 262: ...262 Clarus 590 GC User s Guide ...
Page 263: ...7 Setting up a Typical Analysis ...
Page 264: ...264 Clarus 590 GC User s Guide ...
Page 279: ...Index ...
Page 280: ...280 Clarus 590 GC User s Guide ...